Crown and bridge frameworks, telescopic and conical work, implant bars, implant superstructures and cast partial denture frameworks in cobalt-chrome, 5-axis milled or laser-melted. At a glance: which alloy is approved for which indication.
In short: DD Bio CW from Dental Direkt is milled and approved for crowns and bridges of any span, primary and secondary components, telescopic and conical crowns, implant bars and implant superstructures. In the selective laser melting (SLM) process, we use Starbond Easy Powder from Scheftner and Mediloy S-Co from BEGO; Mediloy S-Co is additionally approved for cast partial denture frameworks and bars. Production runs on our own lines, separate from titanium and zirconia.
Which alloy for which use?
Select the planned case to see which alloy the manufacturer approves for it. “No manufacturer statement” means the instructions for use do not explicitly name that application.
First choiceSuitableNot suitableNo manufacturer statement
Crown and bridge up to 3 units
Framework for veneering
First choice
DD Bio CWStarbond Easy PowderMediloy S-Co
Long-span bridge
from 4 units
First choice
DD Bio CWMediloy S-Co
Also suitable
Starbond Easy Powder
Telescopic and conical work
Primary and secondary components
First choice
DD Bio CW
Also suitable
Starbond Easy Powder
No manufacturer statement: Mediloy S-Co
Implant bar
on implants
First choice
DD Bio CWMediloy S-Co
No manufacturer statement: Starbond Easy Powder
Implant superstructure
screw-retained or cemented
First choice
DD Bio CW
Also suitable
Mediloy S-Co
No manufacturer statement: Starbond Easy Powder
Cast partial denture framework
Clasp and combination prosthetics
First choice
Mediloy S-Co
Also suitable
Starbond Easy Powder
No manufacturer statement: DD Bio CW
Inlay and onlay
metal
First choice
DD Bio CW
No manufacturer statement: Starbond Easy Powder, Mediloy S-Co
Cobalt-chrome in the selective laser melting (SLM) process
The alloys in detail
Bright blue means manufacturer-approved, light blue means approved with limitations, and pale means the instructions for use do not provide for that application. The bar shows the yield strength according to the manufacturer (scale up to 1200 MPa).
DD Bio CW
milledType 4Modulus of elasticity 250 GPa
Milled CoCr alloy from Dental Direkt: crowns and bridges of any span, telescopic and conical crowns, implant bars and implant superstructures.
CoCr powder from Scheftner for selective laser melting: fixed and removable dental prostheses, veneered with ceramic or resin.
Crown and bridge up to 3 unitsFirst choice
Long-span bridgeSuitable
Telescopic and conical workSuitable
Implant barNo manufacturer statement
Implant superstructureNo manufacturer statement
Cast partial denture frameworkSuitable
Inlay and onlayNo manufacturer statement
Mediloy S-Co
SLMType 5Modulus of elasticity 228 GPa
CoCr powder from BEGO for selective laser melting: crowns, bridges, cast partial denture frameworks, bars and secondary bars, including screw-retained implant work.
Crown and bridge up to 3 unitsFirst choice
Long-span bridgeFirst choice
Telescopic and conical workNo manufacturer statement
Implant barFirst choice
Implant superstructureSuitable
Cast partial denture frameworkFirst choice
Inlay and onlayNo manufacturer statement
Not for: ceramic firing on abutments and screw-retained implant bridges
Strength comparison
Laser-melted alloys achieve a higher yield strength than milled DD Bio CW. The milling alloy is softer, easier to finish and explicitly approved for telescopic and conical work.
Yield strength (Rp0.2) according to the manufacturer’s instructions for use; for Mediloy S-Co, the lighter area is between the value after ceramic firing (755 MPa) and after stress-relief annealing (1000 MPa). Minimum wall thicknesses and connector cross-sections are given in the instructions for use.
Datasheets and instructions for use
Instructions for use, safety data sheet and, where available, declaration of conformity and evidence of biocompatibility. Tap the name to open.
DD Bio CWmilled CoCr, Dental Direkt
Type 4 to ISO 22674, modulus of elasticity 250 GPa, hardness 285 HV10, CTE 14.4, yield strength 400 MPa, tensile strength 680 MPa. Can be veneered with ceramic or resin; oxidation firing is not mandatory.
Starbond Easy PowderCoCr in the selective laser melting process, Scheftner
Type 5 to ISO 22674, nickel-free and beryllium-free, modulus of elasticity 225 GPa, hardness 425 HV10, CTE 14.5, yield strength 760 MPa. Stress-relief firing from 4 units; the instructions for use refer generally to fixed and removable dental prostheses.
Mediloy S-CoCoCr in the selective laser melting process, BEGO
Type 5 to ISO 22674, modulus of elasticity 228 GPa, hardness 470 HV10, CTE 14.0, yield strength 1000 MPa after stress-relief annealing (755 MPa after ceramic firing).
Both have their place. Milled DD Bio CW is approved for telescopic and conical crowns, implant bars and implant superstructures. Laser-melted alloys are stronger and suitable for frameworks, cast partial denture frameworks and larger constructions. If nothing is specified, we choose the appropriate process based on the indication.
Which CoCr alloys does CADmill use?
DD Bio CW from Dental Direkt for milling, plus Starbond Easy Powder from Scheftner and Mediloy S-Co from BEGO in the selective laser melting process. Instructions for use and safety data sheets are available on this page.
Can I veneer CoCr frameworks with ceramic?
Yes, with a veneering ceramic whose CTE matches the alloy. The values are given in the instructions for use. For Mediloy S-Co, ceramic firing on abutments and screw-retained implant bridges is not approved.
What if there is a CoCr intolerance?
In that case, titanium is the alternative; see the Titanium page. If an allergy is suspected, the manufacturers recommend an allergy test in advance.
Your privacy
With your consent we use Google Ads and Meta (Facebook) to measure the success of our advertising. Technically necessary functions and an anonymous, cookie-free visit count always run. You can change your choice at any time under “Cookie settings”. Privacy policy